Automated sequential derivatization for Gas Chromatography-[Orbitrap] Mass Spectrometry based metabolite profiling of human blood-based samples
- Organism: Homo sapiens
- Instrument: Q Exactive GC Orbitrap
- SpikeIn:
No
- Keywords:
Automation, derivatization, Thermo Scientific™ TriPlus™ RSH, metabolite profiling, gas chromatography–mass spectrometry (GC-MS)
-
Lab head: Elliott J. Price
Submitter: Elliott J. Price
Many small molecules require derivatization to increase their volatility and to be amenable for gas chromatographic (GC) separation. Derivatization is usually time-consuming, and typical batch-wise procedures increase sample variability. Sequential automation of derivatization via robotic liquid handling enables the overlapping of sample preparation and analysis, maximizing time efficiency and minimizing variability. Herein, a protocol for the fully automated, two-stage derivatization of human blood-based samples in-line with GC–[Orbitrap] mass spectrometry (MS)-based metabolomics is described. The protocol delivers a sample-to-sample runtime of 31 minutes, being suitable for better throughput routine metabolomic analysis.
Created on 8/22/24, 11:21 PM