Error configuring new instrument in Panorama Taher Elgierari  2020-03-26 14:05:14
 

just configured a new instrument in Panorama, that went fine, but when configuring the AutoQC I kept getting the an error saying the folder i entered is not a Panorama folder, i confirmed the folder directory and name matches what in Panorama.
thanks.
taher

 
 
Vagisha Sharma responded:  2020-03-26 14:58:12

Hi Taher,

"Thermo_Orbitrap-2/PepCalMix in Matrix" that is entered in the folder field in your AutoQCLoader screenshot is not a Panorama QC type folder. If you navigate to that folder in a browser you will not see the "QC Summary" and "QC Plots" panel that you expect to see in a QC folder.

Here is the documentation on folder creation in Panorama: https://panoramaweb.org/wiki/home/page.view?name=Create a folder in Panorama.
In the final step of folder creation please select "QC" as the Panorama folder type.

It is probably easiest to delete the folder and create a new one of the correct type. To delete a folder, first navigate to the folder, and then go the the "Admin" menu (gear icon in the top right corner) and select "Folder" > "Management". Confirm that the folder you want to delete is selected in the "Folder Tree" tab and click the "Delete" link.

Let me know if you have any questions.

Best regards,
Vagisha