System Suitability

System Suitability
System taken down to clean quadrupole.2022-04-01 00:00:00 <924>Figure 2 System Suitability2024-04-09 15:12:022024-04-09 15:12:02
Replaced Pump S buffer.2022-03-21 22:41:48 <923>Figure 2 System Suitability2024-04-09 15:12:022024-04-09 15:12:02
LC issue - speculate related to Pump S buffer?2022-03-21 01:25:57 <923>Figure 2 System Suitability2024-04-09 15:12:022024-04-09 15:12:02
New operator. 1 ug injection of plasma-derived particles.2022-03-13 10:11:56 Technician ChangeFigure 2 System Suitability2024-04-09 15:12:022024-04-09 15:12:02
Swapped to PepSep column.2022-03-13 10:11:56 <925>Figure 2 System Suitability2024-04-09 15:12:022024-04-09 15:12:02
Clean quadrupole and hexapole cartridge2021-05-29 00:00:00 Instrumentation ChangeImpact-II weekly QC2021-05-31 08:24:512021-05-31 08:24:51
Cleaning of glass capillary2021-05-28 00:00:00 Instrumentation ChangeImpact-II weekly QC2021-05-28 21:52:322021-05-28 21:52:32
new PicoTip emitter (New Objective, Stock #: FS360-50-15-D-20, DOM: 12-04-2018, Coating: 1P-4P, Tip: 15+/-1um)2021-04-15 00:00:00 Instrumentation ChangeImpact-II weekly QC2021-04-16 06:55:092021-04-16 06:55:09
new PicoTip emitter (New Objective, Stock #: FS360-50-15-D-20, DOM: 12-04-2018, Coating: 1P-4P, Tip: 15+/-1um)2020-10-27 00:00:00 Instrumentation ChangeImpact-II weekly QC2020-11-06 08:17:572020-11-06 08:17:57
Binary pump repair (Halani); new trapping column Symmetry C18 (PN: 186006526; SN: 03033707416509); new nanoT; new separating column Peptide BEH C18 (PN: 186003545; SN: 02833626616217); new ZDV connection (Waters Nr. 289000439)2020-10-01 00:00:00 Instrumentation ChangeImpact-II weekly QC2020-10-03 14:13:492020-10-06 23:18:41
Coll.RF 762 -> 1975; F3L5 -10 -> 0; Corr.Lens 6100 -> 6300; Push pull 1860 -> 1850; Corr.Fill 60 -> 64; Corr.Extr. 975 -> 995; F3L4 3.9 -> 3.02020-09-23 00:00:00 Instrumentation ChangeImpact-II weekly QC2020-09-23 16:50:152020-09-24 15:00:15
Clean Hexapole cartridge, New default instrument parameter tuning: F3L5 0.0 -> -10.0V, F2 350 -> 300, Hex 400 -> 300, Coll.RF 2000 -> 762, Push Pull 1850 -> 1860, Corr.Extr. 995 -> 975, Corr.Lens 6300 -> 6100, Corr.Fill 64.4 -> 60.0, F1L2: 32.4 -> 32.0, Decel. 878.1 -> 877.0, F3L4 3.0 -> 3.9 , F2L1 -16.8 -> -15.8, TOFdetect.Autotune 2287 -> 2267; change picoemitter tip2020-09-17 00:00:00 Instrumentation ChangeImpact-II weekly QC2020-09-21 08:32:432020-09-21 08:34:10
Instrument tuning (voltages): Pulser Push Pull 1850 -> 1850, F3L3 -77.8 -> -86.0, F2L3 27.8 -> 25.7, Reflector 2495 -> 2495, Corr.Fill 62.8 -> 64.4, Corr.Extr. 995 -> 995, Corr.Lens 6400 -> 6300, TOFdetect.Autotune 2237.8 -> 2287, F1 300 - > 150, F2 350 -> 200, H 400 -> 3002020-09-02 00:00:00 Instrumentation ChangeImpact-II weekly QC2020-09-02 09:50:152020-09-02 13:46:30
Cleaning of glass capillary2020-08-28 00:00:00 Instrumentation ChangeImpact-II weekly QC2020-08-28 14:03:492020-08-28 14:03:49
PicoTip emitter (New Objective, Stock #: FS360-50-15-D-20, DOM: 12-04-2018, Coating: 1P-4P, Tip: 15+/-1um)2020-08-27 00:00:00 Instrumentation ChangeImpact-II weekly QC2020-08-28 14:02:482020-08-28 14:02:48
Repair of binary pump (BSM), change of some solvent lines; annual PM SM and BSM;2020-08-26 00:00:00 Instrumentation ChangeImpact-II weekly QC2020-08-28 14:02:122020-08-28 14:02:12
PicoTip emitter (New Objective, Stock #: FS360-50-15-D-20, DOM: 12-04-2018, Coating: 1P-4P, Tip: 15+/-1um)2020-07-15 23:00:00 Instrumentation ChangeImpact-II weekly QC2020-07-16 09:52:212020-07-30 13:00:41
PicoTip emitter (New Objective, Stock #: FS360-50-15-D-20, DOM: 12-04-2018, Coating: 1P-4P, Tip: 15+/-1um)2020-06-12 00:00:00 Instrumentation ChangeImpact-II weekly QC2020-06-17 12:09:362020-07-25 16:16:35
Glass capillary (Bruker), Serial #: 11483-1-28 (Mfg part 38821, RGP T 6.48- ,5/Date 20-Dec-19)2020-06-12 00:00:00 Instrumentation ChangeImpact-II weekly QC2020-06-17 12:10:122020-07-25 16:16:47
Instrument tuning (voltages)2020-06-12 00:00:00 Instrumentation ChangeImpact-II weekly QC2020-06-17 12:10:372020-06-17 12:10:37
Trap column (Waters Symmetry C18, 10k 2G V/V, 100A, 5 um, 180um x 20 mm, Product #: 186006526, Serial #: 03033704716505)2020-06-10 00:00:00 Instrumentation ChangeImpact-II weekly QC2020-06-17 12:58:362020-07-25 16:14:22
Separation column (Waters BEH C18 10k psi, 130A, 1.7 um, 75 um x 250 mm, Product #: 186003545, Serial #: 02833704026201)2020-06-10 00:00:00 Instrumentation ChangeImpact-II weekly QC2020-06-17 12:59:112020-07-25 16:11:46
Clean Hexapole cartridge2020-06-10 00:00:00 Instrumentation ChangeImpact-II weekly QC2020-06-17 13:00:222020-06-17 13:00:22
Clean Quadrupole2020-06-10 00:00:00 Instrumentation ChangeImpact-II weekly QC2020-06-17 13:01:062020-06-17 13:01:06
Clean Collision cell2020-06-10 00:00:00 Instrumentation ChangeImpact-II weekly QC2020-06-17 13:01:242020-06-17 13:01:24
0.1% Formic acid (ThermoFisher BP1215-500, UN1779) in water (ThermoFisher Optima-LCMS, W6-4, Lot 202363)2020-06-10 00:00:00 Reagent ChangeImpact-II weekly QC2020-07-25 16:03:432020-07-25 16:06:37
Acetonitrile (ThermoFisher Optima-LCMS, A955-4, Lot 191742)2020-06-10 00:00:00 Reagent ChangeImpact-II weekly QC2020-07-25 16:06:142020-07-25 16:06:51
ESI-L Low concentration tuning mix (Agilent Technologies,Part #: 01969-85000, Lot #: 0006457849, expires 30-Apr-2021)2020-06-10 00:00:00 Reagent ChangeImpact-II weekly QC2020-07-25 16:09:182020-07-25 16:09:18
Sample vial empty. Peak areas go down, no change in retention times2018-10-22 13:12:27 <325>Lindsay's Real World System Suitability2018-10-25 16:27:322018-10-25 16:30:42
Instrument calibrated2018-10-12 09:32:10 Instrumentation ChangeLindsay's Real World System Suitability2018-10-25 16:12:532022-07-14 12:36:08
LKP got on the instrument2018-10-11 00:00:00 Technician ChangeLindsay's Real World System Suitability2018-10-25 09:23:512018-10-25 09:23:51
Column clogged2018-08-21 16:20:00 <324>Lindsay's Real World System Suitability2018-10-25 09:24:102018-12-20 00:13:34
Column clogged2018-08-16 21:15:13 <324>Lindsay's Real World System Suitability2018-10-25 16:14:202019-03-19 22:26:33
LKP got on the instrument2018-08-16 00:00:00 Technician ChangeLindsay's Real World System Suitability2018-10-25 09:24:282018-10-25 09:24:28
Instrument calibrated2018-08-16 00:00:00 Instrumentation ChangeLindsay's Real World System Suitability2018-10-25 16:12:082018-10-25 16:12:08
LKP got on the instrument2018-01-12 00:00:00 Technician ChangeLindsay's Real World System Suitability2018-10-25 09:24:472018-10-25 09:24:47
LKP got on the instrument2017-11-02 19:16:00 Technician ChangeLindsay's Real World System Suitability2018-10-25 16:18:432018-10-25 16:20:15
Cory ran QC samples2013-08-22 02:00:00 Technician ChangeSystem Suitability2016-06-04 15:40:362016-06-04 15:40:36
Replaced column2013-08-21 06:00:00 Instrumentation ChangeSystem Suitability2016-02-04 14:00:562016-03-08 10:00:37
New reagent lot2013-08-15 02:00:00 Reagent ChangeSystem Suitability2016-06-04 15:41:232016-06-04 15:41:23
Power loss2013-08-12 02:00:00 Instrumentation ChangeSystem Suitability2016-06-04 15:40:582016-06-04 15:40:58
Empty run SSS_run_1392011-09-23 09:09:56 <237>Site 542017-06-27 14:46:282017-06-27 14:56:01
MS cleaning2011-09-11 08:08:00 Instrumentation ChangeSite 56A2017-06-27 14:53:222017-06-27 14:53:22