The samples were prepped as EvoSep tips in 4 sets, general tip prep protocol is attached. The attached .csv's contain the run order, methods, output directories, etc.
Test plate analysis on Thermo Orbitrap Exploris instrument issue log:
Set 1:
18 May 2021 - 20 May 2021
no failures
Set 2:
20 May 2021 - 21 May 2021
no failures
Set 3:
21 May 2021 - 22 May 2021
no failures
Set 4:
22 May 2021 @ 11:48 PM, over-pressurization failure
Fixed + restarted @ 12:18 PM on 23 May 2021
23 May 2021 @ 11:15 PM overpressurization failure
24 May 2021 @ 1:15 AM power failure
Fixed (needle tip cleaned) + restarted @ 12:52 PM on 24 May 2021