Run order and notes

Test Plate Experiment

The samples were prepped as EvoSep tips in 4 sets, general tip prep protocol is attached. The attached .csv's contain the run order, methods, output directories, etc. 

 

Test plate analysis on Thermo Orbitrap Exploris instrument issue log:

  • Set 1: 

    • 18 May 2021 - 20 May 2021

    • no failures

  • Set 2: 

    • 20 May 2021 - 21 May 2021

    • no failures

  • Set 3: 

    • 21 May 2021 - 22 May 2021

    • no failures

  • Set 4: 

    • 22 May 2021 @ 11:48 PM, over-pressurization failure

      • Fixed + restarted @ 12:18 PM on 23 May 2021

    • 23 May 2021 @ 11:15 PM overpressurization failure

    • 24 May 2021 @ 1:15 AM power failure

      • Fixed (needle tip cleaned) + restarted @ 12:52 PM on 24 May 2021

 

 

  Attached Files  
   
 18May2021-testplate-set1.csv
 18May2021-testplate-set2.csv
 18May2021-testplate-set3.csv
 18May2021-testplate-set4new.csv
 General EvoSep Tip Prep SOP.pdf